In a composite of various activities, the normal knee moves through a range of flexion of up to 155 deg, while at all angles there are normal ranges of “laxity,” both in an anterioposterior direction and in internal-external rotation. The ideal goal is that after a total knee replacement (TKR), the knee moves in a similar way, hence, providing the same amount of stability and freedom of motion. Our goal was to device a standardized testing method for evaluating proposed new TKR designs or existing designs. We developed a desktop knee machine, which subjects knees to combinations of forces and moments at a range of flexion angles. The proposed test method would compare the laxity motions of a given design model with the data from normal knee specimens. The TKRs were designed in the computer and then SLA models were made for testing. Computer analysis used RAPIDFORM software to calculate the laxities. In order to specify the testing method, in particular, the compressive, shear, and torque loads, we tested three different knee models with a range of loads. The magnitudes of the loads were proportionately less than in vivo due to the limitations of SLA models. When testing normal knee specimens, there were no frictional effects due to the exceedingly low coefficient of friction at the joint surfaces. However, in metal-plastic TKRs, the friction can affect the laxity and, hence, the kinematics, considerably. Hence this behavior had to be reproduced in our test method. The conclusion from our experiments was that testing should be carried out at a minimum of two compressive loading conditions, one representing low shear/torque ratios and the other, high shear/torque ratios, in order to obtain a realistic representation of the behavior of new TKR designs.
Skip Nav Destination
Article navigation
Design Of Medical Devices Conference Abstracts
Laboratory Evaluation of Total Knee Replacements (TKRs) to Restore Normal Function
Peter S. Walker,
Peter S. Walker
Department of Orthopaedic Surgery, Laboratory for Minimally-Invasive Surgery,
New York University
Search for other works by this author on:
Gokce Yildirim
Gokce Yildirim
MIS Laboratory,
NYU Medical
Search for other works by this author on:
Luis Vasquez
MIS Laboratory,
NYU Medical
Peter S. Walker
Department of Orthopaedic Surgery, Laboratory for Minimally-Invasive Surgery,
New York University
Gokce Yildirim
MIS Laboratory,
NYU Medical
J. Med. Devices. Jun 2010, 4(2): 027508 (1 pages)
Published Online: August 9, 2010
Article history
Online:
August 9, 2010
Published:
August 9, 2010
Citation
Vasquez, L., Walker, P. S., and Yildirim, G. (August 9, 2010). "Laboratory Evaluation of Total Knee Replacements (TKRs) to Restore Normal Function." ASME. J. Med. Devices. June 2010; 4(2): 027508. https://doi.org/10.1115/1.3442442
Download citation file:
566
Views
Get Email Alerts
Cited By
Validation of a Repurposed pH Monitoring Capsule for Cecal Applications Using Novel Synthetic Cecal Contents
J. Med. Devices (March 2025)
Related Articles
Design Concept for a Total Knee Replacement with Condylar Guiding Features
J. Med. Devices (June,2011)
Design of a Dynamic Stabilization Spine Implant
J. Med. Devices (June,2009)
Design and Preliminary Field Validation of a Fully Passive Prosthetic Knee Mechanism for Users With Transfemoral Amputation in India
J. Mechanisms Robotics (June,2018)
Numerical Study of Shear-Induced Thrombus Formation Over Aterial Stent Struts
J. Med. Devices (June,2009)
Related Proceedings Papers
Related Chapters
On the Measurement of Three-Dimensional Taper Moments Due to Friction and Contact Load in Total Hip Replacement
Modularity and Tapers in Total Joint Replacement Devices
Backlash
Design and Application of the Worm Gear
Understanding the Problem
Design and Application of the Worm Gear